Specifications
Agilent 81101A Specifications
Width
Can be entered as absolute width, duty cycle or trailing edge delay.
Duty cycle values from 0.1% to 95% can be entered directly. For values
>95 % press
SHIFT
and use the Modify knob. Note that pulses may be
deteriorated or skipped due to the inaccuracy of period and width. Hence
for large values, it is better to select complement and enter 100 minus the
required duty cycle value.
Delay
Measured between trigger output and main output. Can be entered as
absolute delay, phase
°
or % of period.
Width
Agilent 81101A
Range:
10 ns to 999.5 s
(max value: period –10 ns)
Accuracy:
±
5 %
±
250 ps
Duty cycle:
0.1% to 95% (depends on period and width;
overprogrammable to 99%)
Delay
Agilent 81101A
Fixed delay from
TRIGGER OUT:
17.0 ns typical
Additional variable range:
0.00 ns to 999.5 s
(max value: period –20 ns)
Accuracy:
±
5 %
±
1 ns