Programming Reference
SCPI Instrument Command List
Value coupling
Range coupling
Offset
Description
Use this command to program the amplitude current of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of current, you first
have to execute the
[:SOURce]:HOLD CURRent
command to enable the
[:SOUR
CE
]:CURR
ENT
subsystem.
The available current range is limited by the combination of:
•
Specified Voltage limits
•
Actual OUTPUT Impedance setting
:OUTPut:IMPedance
•
Actual Expected Load impedance setting:
:OUTPut:IMPedance:EXTernal
Example
To program the amplitude current of the output signal:
Command
:CURR[1]:OFFSet
Long
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:OFFSet
Form
Set & Query
Parameter
Numeric
Parameter suffix
A
with engineering prefixes.
*RST value
0.0
µ
A (50
Ω
into 50
Ω
)
Value coupling
Amplitude = High – Low
Offset =
High – Low
2
:HOLD CURR
:CURR 75MA
Enable CURRENT subsystem
Set OUTPUT amplitude to 75 mA
Amplitude = High – Low
Offset =
High – Low
2