93
Specifications
Agilent 81101A Specifications
Timing Specifications
Period
Period can also be entered as frequency.
There are 2 period generation sources available:
•
startable oscillator (variable frequency oscillator VFO)
•
high-accuracy frequency generator (PLL)
Glitch-free timing changes
You can sweep your timing values without danger of spurious pulses or
drop outs that could cause measurement errors. This applies to
continuous mode with timing values < 100 ms (frequency: < 10 Hz), and
consecutive values between one-half and twice the previous value.
Period
Agilent 81101A
Range:
20 ns to 999.5 s
Resolution:
3.5 digits, 5 ps best case for VFO
4 digits, 1 ps best case for PLL
Accuracy:
PLL: ±0.01%
VFO: ±5%
Repeatability:
typically 4 times better than accuracy
RMS-jitter:
PLL: 0.001% + 15 ps
VFO: 0.01% + 15 ps
Frequency range:
1.00 mHz to 50 MHz