67
Programming Reference
SCPI Instrument Command List
Command
:PULS:TRAN[1]:TRA:AUTO
Long
:[SOURce]:PULSe:TRANsition[1]:TRAiling:AUTO
Form
Set & Query
Parameter
ON|OFF|ONCE
*RST value
ON
Description
Use this command to set/read the automatic coupling of the pulse trailing
edge transition time to the leading edge transition time.
Example
To set leading and trailing edges independently:
Command
:PULS:TRIG[1]:VOLT
Long
[:SOURce]:PULSe:TRIGger[1]:VOLTage[:LEVel][:IMMediate]
[:AMPlitude]
Form
Set & Query
Parameter
TTL | ECL
*RST value
TTL
Description
Use this command to set/read the output levels at the TRIGGER OUT
connector.
ON
The trailing edge transition time is automatically
set to the same value as the leading edge, and is
updated automatically each time the leading edge
transition time changes.
OFF
The trailing edge transition time is independently
programmable.
ONCE
The trailing edge transition time is set ONCE to
the same value as the leading edge.
:PULS:TRAN 6NS
:PULS:TRAN:TRA:AUTO OFF
:PULS:TRAN:TRA 15NS
Set OUTPUT leading edge to 6 ns
Enable independent setting of trailing
edge
Set OUTPUT trailing edge to 15 ns
Summary of Contents for 81101A
Page 1: ...S S1 Agilent 81101A 50 MHz Pulse Generator Reference Guide ...
Page 88: ...88 Programming Reference SCPI Instrument Command List ...
Page 112: ...112 Specifications Pulse Parameter Definitions ...
Page 116: ...116 Index ...
Page 118: ...Copyright Agilent Technologies 1998 2000 Edition E0300 Printed in Germany 81101 91021 ...