8
CI Maintenance
To Remove the CI Source Filament
238
7000/7010 Series TQ GC/MS Operating Manual
To Remove the CI Source Filament
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Tweezers (8710-2460)
Procedure
1
Vent the MS. (See
2
Open the front analyzer chamber. (See
3
“To Remove the 7000 Series CI Source”
4
Remove the screw holding the filament to the CI source body. (See
5
Slide the filament off the CI source assembly. (See
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
The analyzer, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.