7
General Maintenance
To Remove the EI XTR Source
7000/7010 Series TQ GC/MS Operating Manual
163
3
Disconnect wiring from the EI XTR. Do not bend the wires any more than
necessary. (See
“To Connect or Disconnect Wiring to the EI XTR Source”
4
Remove the thumbscrews that hold the ion source in place. (See
.)
5
Pull the ion source out of the source radiator.
Figure 50.
Removing the EI XTR source
Ceramic source board
Ion source
Thumbscrews
Source heater and temperature sensor wires
Source radiator