Operating in CI Mode
To Perform a CI Autotune
7000/7010 Series TQ GC/MS Operating Manual
143
8
Select the
Tune from default settings
check box if you are restarting the
system after a system vent, major servicing, or a power outage. If you clear
the
Tune from default settings
box, the Autotune process starts using
previous tune values.
9
Select the
Save tune file when done
check box to save the new tune
parameters generated by the autotune. Do not select this item if you want to
review the autotune report before saving the newly generated tune
parameters.
10
Select the
Print autotune report
check box to automatically print a tune
report.
11
Click
Autotune
to start the autotune. The Status line displays the current step
in the autotune process, and the plot of the tuned parameter for that step is
shown in the top graph. If specified, at the completion of the autotune a Tune
Report is printed.
To stop the autotune before it completes the autotune parameter selection,
click
Abort Autotune
. The parameters from the last successful autotune are
used.
12
Review the Tune Report. If the results are acceptable and you did not select
the
Save tune file when done
check box, save the autotune by clicking the
Files and Report tab
, then click
Save
.