7
General Maintenance
To Remove the EI HES
160
7000/7010 Series TQ GC/MS Operating Manual
5
Using the source finger grip, pull the ion source out of the source radiator.
The source contacts have spring loaded pins so some force must be applied
to pull the source out. (See
Figure 48.
The EI HES before removal
Ion source
Thumbscrew
Source finger grip
Thumbscrew