CI Maintenance
To Switch from the CI Source to the EI HES
7000/7010 Series TQ GC/MS Operating Manual
225
To Switch from the CI Source to the EI HES
Procedure
1
Vent the MS. (See
on page 124.) The software will prompt
you for the appropriate actions.
2
Open the left side access panel. (See
“To Open Left Side Panel for Access to
3
Open the analyzer chamber door. (See
4
Remove the CI source. (See
“To Remove the 7010 Series CI Source”
5
Place the CI source in the storage container.
6
Remove the CI source radiator. The Radiator should not be removed or
installed with the CI source in place. (See
Radiator from the 7010 Series TQ”
7
Place the CI source radiator in the storage container.
8
Loosen the column nut and remove the column from the GC/MS Interface.
9
Cut the column on the tapered side of the ferrule to remove the ferrule.
10
Install the column into the GC/MS interface extending 4 to 5 mm past the analyzer
end of the transfer line. (See
“To Install a Capillary Column in the GC/MS Interface
Using the Self-Tightening Column Nut”
Column in the GC/MS Interface Using a Standard Column Nut”
11
Remove the EI HES radiator from the storage container.
12
Install the EI HES radiator. (See
“To Install the EI HES Radiator”
13
Remove the EI HES from the storage container.
14
Install the EI HES. (See
15
Pumpdown the MS. (See
CAUTION
Always wear clean gloves while touching the analyzer or any other parts that
go inside the analyzer chamber.
CAUTION
Electrostatic discharges to analyzer components are conducted to the side
board where they can damage sensitive components. Wear a grounded
antistatic wrist strap, and take other antistatic precautions
before
you open the
analyzer chamber. (See