5
Operating in EI Mode
To Set the Mode for the Optional JetClean System
7000/7010 Series TQ GC/MS Operating Manual
113
To Set the Mode for the Optional JetClean
System
1
Click the
MS Parameters
icon in the
Instrument Control
view.
2
Click
Source Cleaning
from the
Triple Quadrupole MS Method Editor
navigation panel.
3
Select a mode from the
Operation
dropdown menu. (See
.)
For detailed information on the JetClean system see the
JetClean Operating
manual
.
Figure 35.
TQ MS Method Editor Source Cleaning panel