CI Maintenance
To Remove the 7000 Series CI Source
220
7000/7010 Series TQ GC/MS Operating Manual
To Remove the 7000 Series CI Source
Materials needed
•
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
•
Tweezers (8710-2460)
Procedure
1
Vent the MS. (See
2
Open the analyzer chamber. (See
“To Open the Front Analyzer Chamber”
3
Disconnect the wires from the ion source. (See
4
Remove the two large thumbscrews that hold the ion source in place.
5
Pull the ion source out of the source radiator, and place it in its storage
container.
WARNING
The analyzers, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Use an antistatic wrist strap, and take other antistatic precautions before
touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not on the wires.