![Agilent Technologies 41501A User Manual Download Page 76](http://html1.mh-extra.com/html/agilent-technologies/41501a/41501a_user-manual_2867862076.webp)
2-36
Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
Installation
Mounting Connectors
NOTE
To Perform an Interlock Circuit Test
To confirm that the interlock circuit is operating properly, perform the following
test.
1. Connect the Intlk terminal of the 4155C/4156C to your interlock circuit.
2. Press System key, then select CALIB/DIAG softkey to display the
SYSTEM: SELF-CALIBRATION/DIAGNOSTICS screen.
3. In the CALIB/DIAG field, select DIAG softkey.
4. In the CATEGORY field, select I/O PERIPH softkey.
5. Move the pointer to
403 (INT.) Interlock LED
.
6. Select EXECUTE softkey.
7. Confirm the following:
•
the LED turns on within 1 second when the interlock circuit is shorted.
•
the LED turns off within 1 second when the interlock circuit is opened.
8. To stop the interlock test, select STOP softkey.
Summary of Contents for 41501A
Page 19: ...Contents Agilent 4155C 4156C User s Guide Vol 1 Edition 11 9 Accessories and Options...
Page 20: ...Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Contents...
Page 21: ...1 Introducing the 4155C 4156C...
Page 41: ...2 Installation...
Page 84: ...2 44 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Installation Maintenance...
Page 85: ...3 Connecting to Network...
Page 129: ...4 File Operations...
Page 152: ...4 24 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 File Operations Backing Up a Diskette...
Page 153: ...5 Print Plot Function...
Page 198: ...5 46 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Print Plot Function Output Examples...
Page 199: ...6 External Keyboard...
Page 203: ...7 Initial Settings...
Page 209: ...8 Specifications...
Page 243: ...9 Accessories and Options...