![Agilent Technologies 41501A User Manual Download Page 218](http://html1.mh-extra.com/html/agilent-technologies/41501a/41501a_user-manual_2867862218.webp)
8-10
Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
Specifications
Hardware
4155C Semiconductor Parameter Analyzer
MPSMU (Medium Power SMU) Specifications
•
Voltage Range, Resolution, and Accuracy
•
Current Range, Resolution, and Accuracy
•
Output terminal/connection
Single triaxial connector, non-Kelvin (no remote sensing)
•
Voltage/Current Compliance (Limiting)
The SMU can limit output voltage or current to prevent damaging the device
under test.
Voltage
0 V to ± 100 V
Current
± 1 pA to ± 100 mA
Compliance accuracy
Same as current (voltage) set accuracy.
•
MPSMU supplemental information
Voltage source output
resistance
typical 0.3
Ω
Voltage measurement input
resistance
≥
10
13
Ω
Current source output
resistance
≥
10
13
Ω
(1 nA range)
Current compliance setting
accuracy
(for opposite polarity)
For 1 nA to 10 nA range:
V/I setting accuracy ± 12 % of range
For 100 nA to 100 mA range:
V/I setting accuracy ± 2.5 % of range
Summary of Contents for 41501A
Page 19: ...Contents Agilent 4155C 4156C User s Guide Vol 1 Edition 11 9 Accessories and Options...
Page 20: ...Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Contents...
Page 21: ...1 Introducing the 4155C 4156C...
Page 41: ...2 Installation...
Page 84: ...2 44 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Installation Maintenance...
Page 85: ...3 Connecting to Network...
Page 129: ...4 File Operations...
Page 152: ...4 24 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 File Operations Backing Up a Diskette...
Page 153: ...5 Print Plot Function...
Page 198: ...5 46 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Print Plot Function Output Examples...
Page 199: ...6 External Keyboard...
Page 203: ...7 Initial Settings...
Page 209: ...8 Specifications...
Page 243: ...9 Accessories and Options...