Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
2-15
Installation
Installing Accessories
Connecting the 4156C and the 16442A/B test fixture
Connect the 4156C and the 16442A/B as shown below.
CAUTION
To prevent electrical shock during use, be sure to connect the interlock cable.
WARNING
There are potentially hazardous voltages of up to ±100 V at the Force, Sense,
and Guard terminals. To prevent electrical shock, do
not
expose these lines.
CAUTION
Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the SMU.
4156C
16442A/B
4. HRSMUs
2. VSUs
3. VMUs
1. Intlk
Force
Sense
4156C Terminals
16442A/B
Terminals
Cable
Intlk
Intlk
3 m or 1.5 m Interlock Cable
VSU
VSU
3 m or 1.5 m Coaxial Cable
VMU
VMU
3 m or 1.5 m Coaxial Cable
HRSMU
SMU1 and SMU2
3 m or 1.5 m Kelvin Triaxial Cable
(for Kelvin connection) or
3 m or 1.5 m Triaxial Cable (for
non-Kelvin connection)
SMU3 and SMU4
SMU5 and SMU6
Summary of Contents for 41501A
Page 19: ...Contents Agilent 4155C 4156C User s Guide Vol 1 Edition 11 9 Accessories and Options...
Page 20: ...Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Contents...
Page 21: ...1 Introducing the 4155C 4156C...
Page 41: ...2 Installation...
Page 84: ...2 44 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Installation Maintenance...
Page 85: ...3 Connecting to Network...
Page 129: ...4 File Operations...
Page 152: ...4 24 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 File Operations Backing Up a Diskette...
Page 153: ...5 Print Plot Function...
Page 198: ...5 46 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Print Plot Function Output Examples...
Page 199: ...6 External Keyboard...
Page 203: ...7 Initial Settings...
Page 209: ...8 Specifications...
Page 243: ...9 Accessories and Options...