![Agilent Technologies 41501A User Manual Download Page 217](http://html1.mh-extra.com/html/agilent-technologies/41501a/41501a_user-manual_2867862217.webp)
Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
8-9
Specifications
Hardware
VMU (Voltage Monitor Unit) Specifications
•
Voltage Range, Resolution, and Accuracy
Grounded measurement mode
Differential measurement mode
•
VMU supplemental information
Input impedance
≥
1 G
Ω
Input leakage current
≤
500 pA (at 0 V)
Measurement noise
typical 0.01 % of range (p-p). when the integration
time is 10 PLC.
Differential mode
measurement noise
typical 0.005 % of range (p-p). when the integration
time is short.
Voltage Range
Meas. Reso.
Meas. Accuracy
a
a. Accuracy:
±
(% of reading value + offset value).
±
2 V
2
μ
V
±
(0.02 % + 200
μ
V)
±
20 V
20
μ
V
±
(0.02 % + 1 mV)
Differential
Voltage
Range
Meas.
Reso.
Meas. Accuracy
a
a. Accuracy:
±
(% of reading value + offset value + common mode error).
Max.
Common
Mode
Voltage
±
0.2 V
0.2
μ
V
±
(0.03 % + 10
μ
V + 0.3
μ
V
×
Vi
b
)
b. Vi is the input voltage of VMU2 in volts.
±
20 V
±
2 V
2
μ
V
±
(0.02 % + 100
μ
V + 3
μ
V
×
Vi
b
)
±
20 V
Summary of Contents for 41501A
Page 19: ...Contents Agilent 4155C 4156C User s Guide Vol 1 Edition 11 9 Accessories and Options...
Page 20: ...Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Contents...
Page 21: ...1 Introducing the 4155C 4156C...
Page 41: ...2 Installation...
Page 84: ...2 44 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Installation Maintenance...
Page 85: ...3 Connecting to Network...
Page 129: ...4 File Operations...
Page 152: ...4 24 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 File Operations Backing Up a Diskette...
Page 153: ...5 Print Plot Function...
Page 198: ...5 46 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Print Plot Function Output Examples...
Page 199: ...6 External Keyboard...
Page 203: ...7 Initial Settings...
Page 209: ...8 Specifications...
Page 243: ...9 Accessories and Options...