Agilent 4155C/4156C User’s Guide Vol.1, Edition 11
8-21
Specifications
Functions
•
VAR1'
Staircase or pulsed sweep synchronized with the VAR1 sweep. Sweep is made
with a user specified ratio and offset value. VAR1' output is calculated as
VAR1' = a
×
VAR1 + b, where "a" is the user-specified ratio and "b" is the
user-specified offset value.
•
CONSTANT
A source unit can be set as a constant voltage or current source depending on the
unit.
•
PULSE
One of SMUs can be set as a pulse source.
Pulse width
0.5 ms to 100 ms, 100
μ
s resolutions.
Pulse period
5 ms to 1 s (
≥
pulse width + 4 ms), 100
μ
s resolutions.
SMU pulse setting accuracy (supplemental information):
Conditions: fixed range measurement except multi channel measurement
Width
0.5 % + 50
μ
s
Period
0.5 % + 100
μ
s
Trigger out delay (for pulsed measurements.)
0 to 32.7 ms with 100
μ
s resolution (< pulse width)
Summary of Contents for 41501A
Page 19: ...Contents Agilent 4155C 4156C User s Guide Vol 1 Edition 11 9 Accessories and Options...
Page 20: ...Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Contents...
Page 21: ...1 Introducing the 4155C 4156C...
Page 41: ...2 Installation...
Page 84: ...2 44 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Installation Maintenance...
Page 85: ...3 Connecting to Network...
Page 129: ...4 File Operations...
Page 152: ...4 24 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 File Operations Backing Up a Diskette...
Page 153: ...5 Print Plot Function...
Page 198: ...5 46 Agilent 4155C 4156C User s Guide Vol 1 Edition 11 Print Plot Function Output Examples...
Page 199: ...6 External Keyboard...
Page 203: ...7 Initial Settings...
Page 209: ...8 Specifications...
Page 243: ...9 Accessories and Options...