84
2000/3000 X-Series Oscilloscopes Service Guide
4
Troubleshooting
To prepare for internal assembly troubleshooting
1
Disconnect any external cables from the front panel.
2
Disconnect the power cord.
3
Remove the bucket assembly following the instructions on
page 105
.
4
Remove the power supply shield following the instructions
on
page 107
.
5
Separate the front and rear decks following the
instructions on
page 110
.
WA R N I N G
HAZARDOUS VOLTAGES EXIST — REMOVE POWER FIRST !
The procedures described in this section are performed with power supplied to the
oscilloscope and with the protective covers removed. Only trained service
personnel who are aware of the hazards involved should perform the procedures.
Whenever possible, perform the procedures with the power cord removed from the
oscilloscope. Read the safety notice at the back of this book before proceeding.
WA R N I N G
HAZARDOUS VOLTAGES EXIST — HIGH VOLTAGE IS PRESENT ON POWER
SUPPLY HEAT SINKS !
The power supply heat sinks of the 2000/3000 X-Series oscilloscopes are at a high
potential. This presents an electric shock hazard. Protect yourself from electric
shock by keeping this area covered or by not coming in contact with the heat sinks
when the power cord is attached to the oscilloscope!
C A U T I O N
REMOVE POWER TO AVOID DAMAGE !
Do not disconnect any cables or remove any assemblies while power is applied to
the oscilloscope, or damage to the oscilloscope can occur.
C A U T I O N
AVOID ESD DAMAGE TO COMPONENTS !
ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. Use
proper ESD precautions when doing any of the procedures in this chapter. As a
minimum, place the oscilloscope on a properly grounded ESD mat and wear a
properly grounded ESD strap.
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Page 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Page 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Page 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...