Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
57
To verify trigger sensitivity
This test verifies the trigger sensitivity. In this test, you will
apply a sine wave to the oscilloscope at the upper
bandwidth limit. You will then decrease the amplitude of the
signal to the specified levels, and check to see if the
oscilloscope is still triggered.
Test limits for:
•
Internal trigger sensitivity on all models:
•
< 10 mV/div: greater of 1 div or 5 mV
pp
•
>= 10 mV/div: 0.6 div
•
External trigger sensitivity on all models:
•
DC to 100 MHz: < 200 mV
pp
•
100 MHz - 200 MHz: < 350 mV
pp
Table 16
Equipment Required to Verify Trigger Sensitivity
Equipment
Critical Specifications
Recommended Model/Part
Signal
Generator
25 MHz, 100 MHz, 350 MHz,
500 MHz, and 1 GHz sine
waves
Agilent N5181A
Power splitter Outputs differ < 0.15 dB
Agilent 11667A
Power Meter
Agilent N1914A
Power Sensor
Agilent E9304A or N8482A
Cable
BNC, Qty 3
Agilent 10503A
Adapter
N (m) to BNC (f), Qty 3
Agilent 1250-0780
Feedthrough
50
Ω
BNC (f) to BNC (m)
Agilent 0960-0301
Summary of Contents for 2000 X
Page 1: ...s1 Agilent InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide ...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Figures ...
Page 14: ...14 2000 3000 X Series Oscilloscopes Service Guide Tables ...
Page 22: ...22 2000 3000 X Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Page 144: ...144 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 156: ...156 2000 3000 X Series Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 160: ...160 2000 3000 X Series Oscilloscopes Service Guide Safety Notices ...