Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
39
c
Press
Type:
; then, turn the Entry knob to select
Average - Full Screen
, and press
Add Measurement
.
5
Read the “current” average voltage value as V1.
6
Use the BNC tee and cables to connect the oscilloscope
calibrator /power supply to both the oscilloscope and the
multimeter (see
Figure 3
).
Figure 3
Setting up Equipment for DC Vertical Gain Accuracy Test
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
bananna adapter