Replacing Assemblies
5
2000/3000 X-Series Oscilloscopes Service Guide
107
To remove the power supply shield
1
Using a TORX T10 driver, locate and remove the screw
securing the power supply shield to the rear deck.
Figure 24
Removing the power supply shield screw
WA R N I N G
Sheet metal parts may have sharp edges. Handle with care to avoid injury.