46
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
5
Use the BNC tee and cables to connect the oscilloscope
calibrator /power supply to both the oscilloscope and the
multimeter (see
Figure 5
).
Figure 5
Setting up Equipment for Dual Cursor Accuracy Test
6
Adjust the output so that the multimeter reading displays
the first Volts/div calibrator setting value in
Table 11
or
Table 12
(depending on the oscilloscope model).
7
Disconnect the multimeter.
8
Wait until the measurement settles.
Oscilloscope
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
BNC (f) to dual
bananna adapter