V
erication
and
Service
Information
P
erforming
a
V
erication
Check
12.
Recall
the
position
values
in
register
1
by
pressing:
4
RECALL
5
A
djust
the
right-hand
knob
until
\RECALL:1 "
is
displayed.
13.
Press
4
ENTER
5 .
The
display
will
change
to:
500:500:500:500
14.
Recall
the
position
values
in
register
2
by
pressing:
4
RECALL
5
A
djust
the
right-hand
knob
until
the
display
reads:
''RECALL:2''
15.
Press
4
ENTER
5 .
The
display
will
change
to:
510:510:510:510
V
erify
the
event
timer
16.
Select
the
Event
Timer
function
by
pressing:
4
AUTO/SCAN
5
The
display
will
show
a
reading
similar
to:
00:01
SCAN5
The
rst
half
of
the
display
is
the
event
timer
and
will
display
the
elapsed
time
of
the
scan,
in
a
minute:second
format.
The
verication
test
is
completed.
5-5
Summary of Contents for 11896A
Page 2: ...User Programming and Service Guide Agilent 11896A Polarization Controller ...
Page 17: ...Contents ...
Page 18: ...1 General Information ...
Page 30: ...General Information Speci cations and Characteristics 1 13 ...
Page 35: ...General Information ...
Page 36: ...2 Installation and Preparation for Use ...
Page 54: ...3 Using the Agilent 11896A Polarization Controller ...
Page 59: ...Rear Panel Features The Agilent 11896A rear panel 1 Power input 2 GPIB connector 3 6 ...
Page 65: ...Using the Agilent 11896A Polarization Controller Using the Agilent 11896A ...
Page 66: ...4 Programming ...
Page 100: ...5 Veri cation and Service Information ...
Page 121: ...Veri cation and Service Information ...
Page 122: ...A Choosing the Scan Rate and Measurement Time ...
Page 126: ...B Measurement Considerations ...
Page 131: ...Measurement Considerations ...
Page 132: ...Index ...