d-lab.2, d-lab.2plus, netlab.2plus: Preinstallation
Operation Site
AgfaPhoto
2005-05-01 / PN 04100_01
11
4.7
Emissions during Operation
Photochemical solutions may discharge gases into the air. Some gases have a very low
odor level. In unfavorable cases this may result in bad odor although the maximum
working site concentration (max. value) has not yet been reached. Bad odors may be
avoided by sufficient room ventilation and/or air conditioning.
4.8 Air
Conditioning
4.8.1
Air Conditioning System
During operation the minilab produces heat (details see
Technical data)
Machines with gas laser:
d-lab.2
Type
8060/260
d-lab.2plus Type
8060/261
appr. 5.500 kcal/h
Machines with solid state laser:
d-lab.2 select
Type 8060/803 and Type 8060/804
d-lab.2 basic
Type 8060/270 and Type 8060/272
d-lab.2 (70mm)
Type 8060/157 + 8060/255
d-lab.2plus select
Type 8060/813 and Type 8060/814
d-lab.2plus basic
Type 8060/271 and Type 8060/273
d-lab.2plus (70mm)
Type 8060/158 + 8060/255
netlab.2plus basic
Type 8060/505
appr. 4.640 kcal/h
Therefore we recommend to install air conditioning, which guarantees for the required
room climate (room temperature and humidity) (see sections
Room ventilation
and
Operating conditions
).
The laser exhaust (only d-lab.2 with gas laser, Type 8060/260, and d-lab.2plus with gas
laser, Type 8060/261) can be extracted separately (see
Connection of an exhaust on
the laser exhaust duct
)
Regular exchange of the air volume in the work rooms is necessary.
Note
The wet section of the Paper Processor is sealed. The exhaust air generated here is
negligible. Hence, the connection of an air exhaust is not necessary in this area.
4.8.2 Room
ventilation
Ventilation should be sufficient so that the maximum working site concentration is not
exceeded.
Observe national regulations!