Operation Site
d-lab.2, d-lab.2plus, netlab.2plus: Preinstallation
8
2005-05-01 / PN 04100_01
AgfaPhoto
4.3
Required Space for Operation, Maintenance, and Service
320 cm
60 cm
110 cm
77 cm
15 cm
87 cm
90 cm
A
B
G_GG129
A Hot air extraction (for laser exhaust = 220 m³/h
(only machines with Gaslaser:
d-lab.2
Type 8060/260
d-lab.2plus Type 8060/261)
B Power
connection
The following free space must be reserved for operation, maintenance, and service:
–
Complete floor space: 4.90 m x 1.82 m, approx. 9 m²
–
Free space around the machine (for operation):
–
Left:
110 cm
–
Right: 60 cm
–
Front:
90 cm
–
Rear:
15 cm
Note
Installation and Service require a distance from the wall of approx. 60 cm.
Note
The power connection for the machine is located on the Paper Processor (B).
Furthermore storage space is required for photo-paper and photo-chemicals. Storage
conditions:
–
All photographic papers must always be stored in a cool and dry place. An ideal
storage temperature is b2 °C and +10 °C. Storage at 20 °C is possible for a
few days without any problems. Storage temperatures above 30 ºC should be avoided
as they will cause visible changes after only a few days.
–
Opened packages have to be stored at a relative humidity of 50-60%.
Loading the paper in paper magazines requires a darkroom or a comparable facility (e.g.
dark-tent).