Page | 79
Aeroqual AQM60 User Guide
RECEIVER
= 0x55 header command used for receiver command
MONITOR
= 0xAA header command used for monitor reply
GET_OFFSET_FACTOR
= 0x01 Get all sensor reading offset correction factor
SET_OFFSET_FACTOR
= 0x02 Set one sensor offset value
GET_SENSOR_VERSION
= 0x05 get service settings command
GET_OPERATION
= 0x06 Set AQM box operation command
SET_OPERATION
= 0x07 Get AQM box operation command
GET_CONFIGURATION
= 0x08 Request configure of AQM box command
SET_CONFIGURATION
= 0x09 Configure AQM box command
GET_FILE_NAME_LIST
= 0x0A Get all file name list command
DOWNLOAD_A_FILE
= 0x0B Download a file from SD card command
DELETE_A_FILE
= 0x0C Delete a file in SD card command
SENSOR_RESISTANCE
= 0x0D Sensor resistance command
ZERO_SCRUBBER_ON
= 0x14 Turn on zero scrubber
ZERO_SCRUBBER_OFF
= 0x15 Turn off zero scrubber
GET_GAIN_FACTORS = 0x16 request span calibration values
SET_GAIN_FACTOR
= 0x17 Set sensor gain factor
SENSOR_PARAM_DOWNLOAD
= 0x18 download sensor parameters
SENSOR_PARAM_UPLOAD
= 0x19 upload sensor parameters
SET_RTC
= 0x1E Set AQM monitor Real Time Clock.
UPDATE_SENSOR_LIST
= 0x21 command to update sensor list in the controller
SET_RESET_TIMER
= 0x2A Set auto reset sensor power timer, if none of sensor replied
GET_RESET_TIMER
= 0x2B Get auto reset sensor power timer, if none of sensor replied
AQM_INFO
= 0xFB AQM monitor information request command.
RESET
= 0xFD reset controller command
**************************************************************************************************
SENSOR_CODE
These values defined a sensor type respectively. Some type of sensors can
be
extended up to 4 more subtypes, such as, O3 can extend to 0x31, 0x32, 0x33.
Some of them can one only extend 1 subtype for reserving. However, every
extended
sensor type need a special program. For detailed sensor code, please refer
"AQM_SensorList_V6.0.aql".
Aeroqual reserve the rights to change the sensor list without notice.
RESERVED
1 byte reserved not been used for information transfer, can be 0x00
RESERVED_2
2 bytes reserved
RESERVED_3
3 bytes reserved
RESERVED_4
4 bytes reserved
CHECKSUM
a data stream's check sum - that makes the command stream
total bytes sum is zero.
SENSOR_STATUS
* 8 bits (b7b6b5b4b3b2b1b0) monitor and sensor status information
SensorStatus * b0 = 1 sensor failure
* b0 = 0 sensor working fine
Summary of Contents for AQM60