Page | 78
Aeroqual AQM60 User Guide
receiver then can send sensor stream packet (31 bytes) one by one, total of SENSOR_LIST_SIZE
pakects:
RECEIVER, AQM_ID, SENSOR_CODE, SENSOR_NAME, FRACTION_DIGIT, MEASURE_RANGE,
CONVERT_FACTOR, CAL_POINT, UNIT_NAME, CHECKSUM
* UPDATE_SENSOR_LIST - 1 byte command to update sensor list in the controller.
* SENSOR_LIST_SIZE - 1 byte the total sensor module numbers of the sensor list
* SENSOR_CODE - 1 byte sensor code
* SENSOR_NAME - 7 bytes sensor name ASCII code (a string)
* FRACTION_DIGIT - 1 byte to specify valid reading of fraction digits
* MEASURE_RANGE - 4 bytes floating point value to specify sensor measurement range.
* CONVERT_FACTOR - 4 bytes floating point value to specify the gas sensor's ppm to mg/m3
conversion factor values.
* CAL_POINT - 4 bytes floating point value to indicate default span calibration point.
* UNIT_NAME - 7 bytes reading unit name of ASCII string
30A. All sensor list sent finished, will get another acknowledge command of update sensor list within 5 sec, 4
bytes:
MONITOR, AQM_ID, UPDATE_SENSOR_LIST, CHECKSUM
31. Reset controller command, 4 bytes:
MONITOR, AQM_ID, RESET, CHECKSUM
* RESET - 1 byte command to reset controller, please refer section 2 for its value
31A. Acknowledge reply command of reset controller, 4 bytes:
SENSOR, AQM_ID, RESET, CHECKSUM
* RESET - 1 byte command to reset controller, please refer section 2 for its value
32. Set monitor auto reset sensor power timer command, 4 bytes, followed by 8 bytes data:
MONITOR, AQM_ID, SET_RESET_TIMER, CHECKSUM
MONITOR, AQM_ID, SET_RESET_TIMER, TIMER, CHECKSUM
* SET_RESET_TIMER - 1 byte command
* TIMER - long integer, 4 bytes.
32A. Acknowledge reply command for SET_RESET_TIMER, 4 bytes:
SENSOR, AQM_ID, SET_RESET_TIMER, CHECKSUM
33. Get monitor auto reset sensor power timer command, 4 bytes:
MONITOR, AQM_ID, GET_RESET_TIMER, CHECKSUM
33A. Reply data of GET_RESET_TIMER command, 8 bytes:
SENSOR, AQM_ID, GET_RESET_TIMER, TIMER, CHECKSUM
*TIMER - long integer, 4 bytes
*************************************************************************************************