OPTION 008 RF PROFILES AND COMPLEX SWEEP
Annex-B-6
Carrier :
Freq.
RF Level :
dBm
Single Modulation Mode
Modulation ENABLED
MHz
Int Std: 10 MHz
Offset: +20.0 dB
FM:
Hz
Int F4:
1.0000 kHz
0
ON
100.000 0000
+0.0
ON
FM
Devn.
Source
Freq : F4
FM
ON/OFF
Select
Source
RF
Level
Carrier
Freq.
LOCAL
Low
Intermod
Φ
M
AM
Wideband
FM
C0863
Fig. B-4 Main signal generator screen with RF offset enabled
RF profiles
If the
[Profiles]
key is selected from the menu shown in Fig. B-2 the display shown in Fig. B-5 is
displayed. But note that the
[Save Setting]
and
[Edit Profile]
keys will only appear if the
instrument is unlocked to Level 1.
Select
Profile
LOCAL
Enable/
Disable
Save
Setting
Edit
Profile
EXIT
RF Profile Menu
Profile Selected:
1 -
Profile: DISABLED
C0865
Fig. B-5 RF profile menu (shown unlocked to Level 1)
Up to 10 profiles may be generated and stored using this facility. The profiles are identified as
Profile 0 to Profile 9. Each profile can have up to 100 frequencies at which the output level of the
signal generator can be adjusted by up to ±40 dB to compensate for the frequency response of an
external device without altering the displayed RF level of the signal generator.