2.6
BUTTON
This button is used to display the frequency measurements of a signal, of power,
of the levels and orders of harmonics.
NOTE:
This button is not functional in the DC mode.
2.6.1
Normal Mode
Successive
presses on
Function
short
Displays:
- The frequency of the signal, the RMS
measurement, and the DC component
- The crest factor CF, the RMS measurement,
and the DC component
long (>2 sec)
- Enters or exits the THD calculation and
display mode
- Displays the THDf, the THDr, and the RMS
value
- The
and
buttons are used to display
each order of harmonic (25 orders, from h01
to h25), with the associated harmonic
distortion (with respect to the fundamental)
and the RMS value of order hxx
-
NOTE:
Order hdC (displayed in the DC and
AC+DC modes) is the DC component; order
h01 is the fundamental
- Starts/Stops the energy metering mode
- Displays the various energy parameters
- The
and
buttons are used to display
the status and energy metering measurement
results pages
2.6.2
Harmonic Order Display
or
+
Successive
presses on
…serve
short
-
First press:
Displays the frequency of the
harmonic order previously selected using the
or
buttons, instead of order hxx
-
Second press:
Restores the display of order
(hxx or hdC)