19
2.5.2
The MAX/MIN Mode + Activation of the HOLD Mode
Successive presses on
Function
short
-
Displays the MAX, AVG, MIN and
PEAK+, AVG, PEAK- values detected
before the
button was pressed
-
When the
button is pressed, the
last value is held on the display
NOTE:
The HOLD function does not interrupt the acquisition of new MAX, MIN, PEAK
values
2.5.3
Access to the True Inrush
™
Mode (
set switch to
)
This button allows measurement of the True Inrush
™
current (starting current, or
overcurrent in steady-state operation) for AC or DC current only (not operational
in AC+DC).
Successive presses on
Function
long (>2 sec)
-
First press:
Enters the True InRush
™
mode
- "Inrh" is displayed for 3s (the
backlighting blinks)
- The triggering threshold is displayed for
5s (the backlighting is steady)
- "------" is displayed and the "A" symbol
flashes (backlighting turns off)
- After detection and acquisition, the
InRush current measurement is
displayed, after the calculations stage
"------" (backlighting off)
NOTE:
The A symbol flashes to indicate
"surveillance" of the signal
-
Second press:
Exits the True InRush
™
mode (returns to simple current
measurement)
short (<2 sec)
Note:
A short press
is functional only if a True
InRush value has been
detected.
- Displays the PEAK+ value of the current
- Displays the PEAK- value of the current
- Displays the RMS True InRush
™
current
NOTE:
The A symbol is displayed during
this sequence