23
3.4
CONFIGURATION
As a safety measure, and to avoid repeated overloads on the inputs of the meter,
configuration operations should only be performed when the meter is
disconnected from all dangerous voltages.
3.4.1
Auto Power Off
The Auto Power Off feature is enabled by default. To disable it, perform the
following:
1.
In the OFF position, hold the
button down while turning the switch
to
until the "full screen" display ends and a beep is emitted. The
symbol is displayed.
2.
When the
button is released, the device is in the voltmeter function
in the normal mode.
3.
To return to Auto Power Off, turn the clamp-on meter OFF and then
back ON again.
3.4.2
Configuring the Current Threshold for True InRush
™
Measurement
To configure the triggering current threshold of the True InRush
™
measurement:
1. In the OFF position, hold the
button down while turning the switch
to
until the "full screen" display ends and a beep is emitted. The
display will indicate the percentage overshoot to apply to the measured
current to determine the measurement triggering threshold.
The value stored by default is 10%, representing 110% of the
established current measured. The possible values are 5%, 10%, 20%,
50%, 70%, 100%, 150%, and 200%.
2. To change the threshold, press the
(yellow) button. The value
flashes; each press on the
(yellow) button displays the next value.
To record the chosen threshold, apply a long press (>2s) on the
(yellow) button. A confirmation beep is emitted.
To exit from the configuration mode, turn the switch to another setting. The
chosen threshold is stored and a double beep is emitted.
NOTE:
The starting (InRush) current measurement triggering threshold is fixed at
1% of the least sensitive range. This value is 1% of 99.99A or 1A. This threshold
is not adjustable.