30
3.13 ADAPTER FUNCTION MEASUREMENT
This function makes it possible to connect any adapter or sensor that converts a
measured electrical or physical quantity into a voltage (either DC or AC). A direct,
immediate reading is obtained without applying a conversion factor.
The measurement mode, AC or DC (the default), must be chosen manually using
the yellow button.
The scale factor for the adapter or sensor must be chosen in advance in the
instrument’s set-up (§3.4.5). The table below indicates the various
adapter/sensor sensitivities that allow a direct reading once the scale factor has
been chosen.
Adapter/Sensor
Sensitivity
Scale Factor to be
Programmed
10 mV/kX (0.01 mV/X)
10 k
100 mV/kX (0.1 mV/X)
100 k
1 mV/X
1
10 mV/X
10
100 mV/X
100
1000 mV/X (1 mV/mX)
1 m
10 mV/mX
10 m
100 mV/mX
100 m
NOTE: The value represented as (X) is valid for any quantity:
humidity (%RH), illumination (lux), speed (m/s), flow, (GPM) etc.
1. Connect the black lead to the
COM
terminal and the red lead to the "
+
"
terminal.
2. Set the switch to
.
3. Select the AC or DC mode.
4. Connect the adapter according to its directions for use.