34
4.2.7 Resistance Measurement
Measurement Range
(1)
0.0 to
599.9
Ω
600 to
5999
Ω
6.00 to
59.99k
Ω
Specified Measurement
Range
1 to 100% of the
measurement range
0 to 100% of the
measurement range
Accuracy
± (1% R +5cts)
Resolution
0.1
Ω
1
Ω
10
Ω
Open-circuit Voltage
≤
3.6V
Measurement Current
550µA
100µA
10µA
Note (1) Above the maximum display value, the display unit indicates "OL”.
The "-" and "+" signs are not managed.
Specific Specifications in MAX/MIN mode
:
•
Accuracy: add 1% R to the values of the table above.
•
Capture of the extreme: approximately 100ms.
4.2.8 Diode test
Measurement Range
0.000 to 3.199V
DC
Specified Measurement Range
1 to 100% of the measurement range
Accuracy
± (1% R + 10cts)
Resolution
0.001V
Measurement Current
0.55mA
Indication: junction reversed or
open-circuit
"OL" is displayed when the measured
voltage >3.199V
Note: The "-" sign is disabled for the diode test function.
4.2.9 Frequency Measurements
4.2.9.1 Voltage
Measurement Range
(1)
5.0 to 599.9Hz
600 to 5999Hz 6.00 to 19.99kHz
Specified Measurement
Range
1 to 100% of the
measurement
range
0 to 100% of the
measurement range
Accuracy
± (0.4% R + 1ct)
Resolution
0.1Hz
1Hz
10Hz