4 SPECIFICATIONS
4.1 REFERENCE CONDITIONS
Quantities of Influence
Reference Conditions
Temperature:
23°C ±2°C
Relative humidity:
45% to 75%
Supply voltage:
9.0V ±0.5V
Frequency range of the applied signal:
45 to 65Hz
Sine wave:
pure
Peak factor of the applied alternating signal:
√2
Position of the conductor in the clamp:
centered
Adjacent conductors:
none
Alternating magnetic field:
none
Electric field:
none
4.2 SPECIFICATIONS UNDER THE REFERENCE CONDITIONS
Accuracy is expressed in ± (x% of the reading (R) + y counts (ct)).
4.2.1 DC Voltage Measurement
Measurement range
0.00 to 59.99V
60.0 to 599.9V
600 to 1000V
(1)
Specified
measurement range
0 to 100% of the measurement range
Accuracy
from 0.00V to 5.99V
±(1% R + 10cts)
from 6.00V to 59.99V
±(1% R +3cts)
±(1% R +3cts)
Resolution
0.01V
0.1V
1V
Input impedance
10M
Ω
Note (1) The display indicates "+OL" above + 2000V and "-OL"
below – 2000V, in REL mode. The "-" and "+" signs are managed.
Above 1000V, a repetitive beep indicates that the voltage being
measured is greater than the safety voltage for which the device is
guaranteed. The display indicates "OL.”