20
3.4 CONFIGURATION
As a safety measure, and to avoid repeated overloads on the inputs of the
device, configuration operations should only be performed when the device is
disconnected from all dangerous voltages.
3.4.1 Configuring the Maximum Resistance for Continuity
To configure the maximum resistance allowed for a continuity:
1. With the switch in the OFF position, hold the
(yellow) button down
while turning the switch to
until the "full screen" display ends and a
beep is emitted. The display will indicate the value below which the
buzzer is activated and the symbol is displayed. The value stored by
default is 40
Ω
. The possible values range between 1
Ω
and 999
Ω
.
2. To change the threshold, press the
(yellow) button. The right-hand
digit flashes; each press on the
(yellow) button increments it. To
shift to the next digit, apply a long press (>2s) to the
(yellow)
button.
When the desired value is displayed, turn the switch to another setting. The
detection threshold chosen is stored and a double beep is emitted.
3.4.2 Auto Power OFF
The Auto Power OFF feature is enabled by default. To disable it, perform the
following:
1. In the OFF position, hold the
button down while turning the switch
to
until the "full screen" display ends and a beep is emitted. The
symbol is displayed.
2. When the
button is released, the device is in the voltmeter function
in the normal mode.
3. To return to Auto Power OFF, turn the clamp-on meter OFF and then
back ON again.
3.4.3 Configuring the Current Threshold for True InRush
®
Measurement
To configure the triggering current threshold of the True InRush
®
measurement:
1. In the OFF position, hold the
button down while turning the switch
to
until the "full screen" display ends and a beep is emitted. The
display will indicate the percentage overshoot to apply to the measured
current to determine the measurement triggering threshold.
The value stored by default is 10%, representing 110% of the
established current measured. The possible values are 5%, 10%, 20%,
50%, 70%, 100%, 150%, and 200%.