4.2.11 Adapter Function Measurement
4.2.11.1 In DC mode
Measurement Range (1)
0.0 to 599.9mV
0.60 to 5.999V
Specified Measurement Range (2)
0 to 100% of the measurement range
Accuracy
1% R ±3cts
Resolution
0.1mV
1mV
Input Impedance
10M
Ω
4.2.11.2 In AC mode
Measurement Range (1)
5.0 to 599.9mV
0.600 to 5.999V
Specified Measurement
Range (2)
1 to 100% of the
measurement range
0 to 100% of the
measurement range
Accuracy
5.0 to 59.9mV
± (1% R + 10cts)
60.0 to 599.9mV
± (1% R +3cts)
1% R ±3cts
Resolution
0.1mV
1mV
Input Impedance
10M
Ω
Note (1) The basic display is
6
000 counts. The position of the decimal count
and the display of multiples (m and k) depend on the programming of
the scale factor.
•
In DC, the display indicates "+OL" above +5999 counts and "-
OL" below -5999 counts. The "-" and "+" signs are managed
(polarity).
•
In AC, the display indicates "OL" above 5999 counts.
Note (2) - The max. bandwidth is 1kHz.
Specific specifications in MAX/MIN mode
(from 10Hz to 1kHz):
•
Accuracy: add 1% R to the values of the table above.
•
Capture of the extreme: approximately 100ms.