ADOS S.R.L. Buccinasco (MI)
417-MTA200I1
Rev. 9
Page 74 of 77
5.8 System
Initialization
The instrument can be reset to the default configuration by mean of the System initialization
function.
To enter the system initialization mode , the user must enter the instrument control mode and use
the
F/
INC
key to select the modality:
S Y S I N I
T
and press the
E
key.
The instrument will ask for password and, if correct, the following message will be displayed:
* I N I T *
Press the
E
key to request to start the sequence and the instrument prepares to receive the
confirmation command:
O K I N I T
?
flashing
Press the
E
key to confirm the selection made and the instrument initializes the parameters and
restarts automatically.
Press the
C
key to cancel the initialization request and return to the instrument control menu.
5.8.1 Switch controlled initialization
Initialization by switch can only be performed at power-on time.
The procedure is the following:
•
DIP switch SW1-1 must be ON prior to power on
•
The display shows the message:
O K I N I T
?
flashing
•
if the switch is set to OFF in less than 5 seconds the initialization is carried out and the message
on display changes to
* I N I T *
flashing
and stays in this condition as long as a key is pressed, thus proceeding with startup sequence.
•
if the switch is left unchanged the instrument proceeds with normal startup sequences