3.3.3 Load cell connection
Connect the cable coming from the load cells or from the junction box to the circular connector
located in the rear of the instrument and indicated as LC.
The load cell must be connected to the instrument with a shielded cable according to the attached
table. To this regard, we recommended the use of our shielded cable (Type CS-4).
The cable shield at A200E side must be grounded. It can be wired to the protective conductor
terminal
The connection must be made in accordance with the color code reported on the load cell (when
using ADOS cells refer to table 167-MTSCACOC).
The A200E includes the standard connection of load cells with 4 wires. For a connection to load
cells equipped with senses (therefore with a 6-wire connection), contact ADOS.
THE CONNECTION CABLE BETWEEN THE LOAD CELLS AND THE INSTRUMENT
MUST FOLLOW A COMPLETELY SEPARATE LAYOUT, TO AVOID UNWANTED
COUPLING THAT CAN REDUCE THE QUALITY OF MEASUREMENT.
Figure 5 - A200E load cell connector
PIN
FUNCTION
ADOS COLOR CODE
A
+ POWER SUPPLY
RED
B
- POWER SUPPLY
BLACK
C
+
SIGNAL GREEN
D
-
SIGNAL WHITE
E N.C.
F N.C.
Cable shield must be connected to a good ground reference plane.