5.2.1 Password
The operator access to configuration, weight calibration and system initialization is controlled by
means of a password.
The password value is “
45700120
”.
The password value is fixed and there are no procedures to modify while in operation. The scope of
the password is not intended for security but only to prevent unwanted access to system functions.
The instrument shows the following message on the auxiliary display:
P A S S W O R
D
and the message characters are substituted with “
*
” when the operator keys in the password.
When a correct access code is fully keyed the selected function menu will appear, otherwise the
instrument will go back to the function selection menu.
5.2.2 Keyboard time out
After 30 seconds of no keyboard activity the instrument automatically generates a
C
key code, so if
the instrument is abandoned while in configuration or calibration it will go back to the operation
menu and than to the operating status.
5.2.3 Keyboard disable function
Selecting parameter “ KBD ENBL” to OFF it is possible to disable all keyboard function. To enable
again the keyboard perform the follow sequence :
hold down the
C
and
E
keys at the same time while the instrument is being turned on
the instrument displays the following message :
R E L K E Y
Release the
C
and
E
keys.
The instrument displays the following message :
P A S S W O R
D
Keys the password then go to configuration and set to ON the parameter “ KBD ENBL ”.