ADOS S.R.L. Buccinasco (MI)
417-MTA200I1
Rev. 9
Page 69 of 77
5.5.1 Zero
calibration
The weighing system must be set with ZERO weight.
After selecting the modality:
Z E R O
press the
E
key.
The instrument prepares to receive the confirmation command and flashes the message on the
auxiliary display.
Press the
E
key again and the instrument starts the zero calibration procedure. The display shows
the waiting time:
W A I T 5
During the waiting time the load cell signal is integrated and it must stay stable.
When the wait time expires instrument performs the zero calibration and displays the following
message for 2 seconds:
D O N E
and returns automatically to the calibration menu.