GB 03438 - Edition 03 - October 09
11/16
A Commissioning
B Product
description
C Settings
D Test
mode
E RS232
mode
F Appendices
Fault
Controller
Gauge
Power LED
& display
Definition
Possible fault
Action
G
Err Lo
«Gauge fault:
signal 0-10V too low
-Measurement stopped-
Gauge error
See Gauges: Fault catalog
Err Id
Identification error of the
gauge
The gauge is not
recognized
Check the gauge
Check the cable
No GAUGE
No gauge connected
No gauge connected
Check the gauge
Check the cable
Appendix 6
Error messages
Led ON
Led OFF