A Commissioning
B Product
description
C Settings
D Test
mode
E RS232
mode
F Appendices
General settings for the
ACS 2000 controller
RS232 settings
Communication speed
Description
This setting is used to adjust the communication speed of the RS232
serial interface.
The speed is expressed in Baud units.
Accessing this setting:
Access:
SET/PROG2/4. BAUD
Modification of this setting does not require the controller to be
restarted.
Factory default setting:
The communication speed is factory set to
9600 bauds
.
Locking the settings
Description
This setting stops anyone modifying the configuration.
It is activated when the LOCK setting is ON.
“Err05” will appear anytime anyone tries to modify a setting.
See appendix 6: error messages).
To unlock the settings, set parameter lock to “
OFF
“
RS232 setting
Selecting the end of line characters for the RS232 transmission
CR (carriage return) = return to start of line only (with a risk of
overwriting the previous data).
CR LF (carriage line Feed) = advance to start of next line
(with no chance of overwriting the previously sent data).
Accessing this setting
Access:
SET/PROG2/7. RS232C
Factory default setting:
The controller is factory set to the
CR
position.
㪧㪩㪦㪞㪄㩷㪉㩷
㪪㪧㩷
㪣㪦㪚㪢㩷
㪈
㪉
Accessing this setting:
Access:
SET/PROG2/5. PLOC
Modification of this setting does not require the controller to be
restarted.
Factory default setting
By default the controller is not locked: LOCK is factory set to “
OFF
“.