Appendix 6
Error messages
Fault
Controller
Gauge
Power LED
& display
Definition
Possible fault
Action
P
Led OFF
No display
Power supply board
1- Check power supply
2- Contact customer Service for repair
Q
Led ON
No display
Display or display control
Contact customer Service for repair
R
Err 01
Eprom fault
Channel control board
1- Press ENTER for 3 seconds
«Setting» appears in flashing then ,
«SET» is displayed. Switch controller
off, then switch on
2- Contact customer Service for repair
Err 02
display
2 seconds
Button activation
«Err Id» Running
Correct «Err Id»
S
Err 03
A/D CPU converter fault
-Measurement stopped-
Main board
1- Switch controller off then
switch on
2- Contact customer Service for repair
T
Err 04
A/D converter fault
-Measurement stopped-
Channel control board
1- Switch controller off then
switch on
2- Contact customer Service for repair
Err 05
display
2 seconds
Modification of a
parameter
«PLOC» function (ON)
running
Deactivate the function «PLOC» (OFF)
E
Err 06
Gauge fault: Eprom or
converter A/D
-Measurement stopped-
Gauge error
See Gauges: Fault catalog
F
Err 07
AHC gauge fault:
Filament B-A or electronic
-Measurement stopped-
Gauge error
See Gauges: Fault catalog
Err 08
display
2 seconds
Starting of the AHC
“Degas” function
-Function not started-
Conditions required not
present
Check the required conditions
Err 09
display
2 seconds
Starting of the A/D
“zero” function
-Function not started-
Conditions required not
present
Check the required conditions
A
Err Hi
Gauge fault:
signal 0-10V too high
-Measurement stopped-
Gauge error
See Gauges: Fault catalog
B
Or
Pressure over the maximum
readable pressure of the
gauge (Over range)
Gauge message
See Gauges: Fault catalog
D
Ur
Pressure below the
maximum readable pressure
of the gauge (Under range)
Gauge message
See gauges: Catalog of faults
Led ON
Led OFF