2
General safety
◆
If condensation has occurred, thoroughly dry out the 761 before startup.
◆
Connecting the REF/FLT port to the reference port of unit under test (UUT) with a small differential pressure range
might result in control problems over time as the environmental temperature changes. Disconnect the ports and
reconnect to resolve control issues.
◆
The vent port should not face the operator during venting.
◆
Do not apply excessive voltage or current to the electrical terminals.
◆
Do not use any adapter other than Additel model 9818. Charge as soon as the battery symbol indicates.
◆
If the calibrator is not working properly, turn it off, remove the battery and contact Additel.
◆
Do not remove the battery while it is charging or when the calibrator is in use.
◆
Before turning off the calibrator, make sure the system pressure is reduced to the atmosphere pressure.