3.4.9.1 Current/voltage measure
1. Calibration points
◆
The default is -30mA, 0mA, 30mA for current measure and -30V, 0V, 30V for voltage measure.
◆
Calibration points can be edited with the respective range as follows if needed:
1) (-30~-18)mA or (-30~-18)V for negative full scale.
2) (-12~12)mA or (-12~12)V for zero.
3) (18~30)mA or (18~30)V for positive full scale.
2. Run
◆
Press / to start.
◆
According to the calibration point prompt by the calibrator, use a more precise, higher-standardized unit generates
signal to the calibrator.
◆
When the measured value is stable, Press / to record and continue, press / to review.
3.4.9.2 Current source
1. Calibration points
◆
The default is 4mA, 20mA that cannot be edited.
2. Run
◆
Press / to start.
◆
The calibrator will provide the current to a standard ampere-meter on schedule. When the reading is stable, input
the reading into the calibrator.
◆
Press / to record and continue and press / to review.
◆
The accepted range of the standard gauge is (0 ~ 30)mA.
3.4.9.3 High/low range
1. Calibration points
◆
If the range is mixed, execute in three-point mode calibration with the low limit, the zero and the high limit as default.
35