2) The name cannot be the same as the existing task.
3) The characters of the input test name cannot exceed 13 spaces.
◆
Output signal: According to the current range of the UUT.
◆
Accuracy.
1) Given: 0.05%, 0.1%, 0.15%, 0.25%, 0.5%, 1.0%, 1.6%.
2) Custom *X.XX%: When the custom range is selected, press F2, and the field can be edited to the desired
accuracy from (0.01 to 100)%FS.
◆
Calibration points.
1) The range is 2~17.
2) Scale will default to linear steps over the chosen range of the UUT.
3) Each test point can be edited individually.
◆
Range of UUT.
1) Cannot exceed the measure the range of the calibrator.
2) The high and low limit cannot be equal.
◆
Cycles: 1~3.
3. Run
◆
Input the information before running including the date (the default is the system date), the ambient temperature (the
default is 25
℃
and the range is 0~50
℃
) and the relative humidity (the default is 40%RH and the range is
0~100%RH).
◆
Pressure Source: Internal Pressure or Pressure Module.
◆
Select the run mode.
1) Automatic: Set the wait time and delay time, and their ranges are both 0~3600. When the pressure is stable, the
wait time will count down to zero, then the task will continue to the next set point. When the run direction is ready
46