5.3.4 Barometer altimeter
1. Connect
◆
Connect the calibrator to the barometer altimeter as shown in Figure 5-4.
2. New
◆
Task name:
1) Enter an alphanumeric name, and press
to input special codes.
2) The name cannot be the same as the existing task.
3) The characters of the input task name cannot exceed
12 spaces.
◆
Range of UUT:
1) Range: 0.1bar~1.2bar.
2) Cannot exceed the output range of the calibrator.
3) The high and low limit cannot be equal.
4) Over pressuring will result in damage to the UUT.
◆
Calibration points:
1) The range is 2-17.
2) Scale the steps in accordance with the UUT.
3) Every point value and tolerance can be edited.
Figure 5-4 Calibrating a barometer altimeter