2) The name cannot be the same as the existing task.
3) The characters of the input task name cannot exceed 12 spaces.
◆
Range of UUT:
1) Range: 0.1bar~1.2bar.
2) Cannot exceed the output range of the calibrator.
3) The high and low limit cannot be equal.
4) Over pressurizing will result in damage to the UUT.
◆
Calibration points:
1) The range is 2-17.
2) Scale the steps in accordance with the linear the UUT for the default.
3) Every point can be edited.
◆
Tolerance:
1) Given: ±(0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 5)mbar.
2) Custom: the custom range is (0-110)mbar.
◆
Resolution.
Given: 1, 0.1, 0.01, 0.001.
◆
Cycles: 1~3, and the default is 2.
3. Run
◆
Input the information before running including the date (the default is the system date), the ambient temperature (the
default is 25
℃
and the range is 0~50
℃
) and the relative humidity (the default is 40%RH and the range is
0~100%RH).
51