DC sensors
DC sensors are used for discrete
current signals, e.g. temperature, pressure, etc .
After selection of DC1 – DC3 or All DC Channels item the sensor properties menu appears.
Sensitivity[mV/unit]
sensitivity value
Offset[mV]
offset value
Unit
the unit selected from the list or the user unit
Position
the angle of sensor (see picture above). Usually used for proximity sensors and GAP
measurement.
The used formula: output value in Unit = (input value in mV - Offset) / Sensitivity.
Tacho
The speed is measured from the tacho channel. The signal contains one or more pulses per one rotation. The
pulses can also trigger the measurement.