© Adam Equipment Company 2010
7.1 SETTING
OF
PIN
•
Display will show
“F2” “Pin”
. This parameter allows setting of a new
password number. The default password is
“0000”
.
•
Press
[Tare]
. The
“Weight”
display will show
"Pin 1"
.
•
Enter the new password number. The
“Unit Weight”
display will show
dashes. Press
[Tare]
.
•
The
“Weight”
display will change to
"Pin 2 "
, Enter the password again
and press
[Tare]
.
•
The display will show
"donE"
to show the new password has been
accepted and will return to the menu. Record the new password in a
secured place.
•
To exit a parameter, press the
[Zero]
key.
7.2 ENABLE
WEIGHING
UNITS
•
Display will show
“F3” “Unit”
. This parameter allows weighing units to be
enabled or disabled. If enabled the weighing units will change when the
[U.Wt./Unit]
key is pressed. See section 6.0.
•
To enter this function press the
[Tare]
key.
•
The display will show the first weighing unit and if it is currently On or
Off. Press the
[U.Wt./Unit]
key to change the setting and then press
the
[Tare]
key to go to the next weighing unit. The weighing units
selected will depend upon the model and the capacity of the scales.
See Section 12.2, Service Parameters.
•
Press the
[Zero]
key to return to the weighing mode.