© Adam Equipment Company 2010
•
To enter the calibration parameter, press the
[Tare]
key when
“F4”
is
displayed. The scale will be calibrated using 2 masses of approximately
1/3Maximum and Maximum. If the scale has been calibrated once the
values will be stored. If this is the first time the scale is calibrated the
user must enter the values for the calibration masses.
•
The display will instruct you to remove any weight from the scale,
“UnloAd”
. Press
[Tare]
.
•
The display will tell you to add the first weight to the scale:
“LoAd 1” “
40” “kiloS”
•
Add the weight shown, wait for stability then press the
[Tare]
key.
•
The display will tell you to add the second weight to the scale:
“LoAd 2 “
100” “kiloS”
•
Remove the weight.
•
The display will show
“SPAn” “PASS”
if the calibration is OK. Or it will
show
“SPAn” “FAIL”
if calibration could not be completed correctly. After
calibration the scale will return to normal operation. If you wish to set
any of the following parameters enter the Service menu again and skip
the first settings by pressing the
[U.Wt./Unit]
key.
12.2.5
F5 –INITIAL ZERO RANGE
•
To enter this parameter, press the
[Tare]
key when
“F5”
is shown.
•
The display will show the current initial zero range.
•
Press the
[U.Wt./Units]
key to change the value and press
[Tare]
to
accept the value.
•
Press
[Zero]
to return to weighing.
12.2.6
F6 -RE-ZERO RANGE
•
To enter this parameter, press the
[Tare]
key when
“F6”
is shown.
•
The display will show the current re-zero range.
•
Press the
[U.Wt./Units]
key to change the value.